Jia Xiaohang, Zhang Qinyuan, Yan Qinglai, et al. The study of the methods for evaluating the matching rate of capsule endoscope pixels and its resolution effectiveness[J]. Opto-Electronic Engineering, 2019, 46(1): 180112. doi: 10.12086/oee.2019.180112
Citation: Jia Xiaohang, Zhang Qinyuan, Yan Qinglai, et al. The study of the methods for evaluating the matching rate of capsule endoscope pixels and its resolution effectiveness[J]. Opto-Electronic Engineering, 2019, 46(1): 180112. doi: 10.12086/oee.2019.180112

The study of the methods for evaluating the matching rate of capsule endoscope pixels and its resolution effectiveness

    Fund Project: Supported by National Key Research and Development Plan Fund (2016YFC0107000)
More Information
  • The evaluation for the matching level between the optical resolution and the pixels number on the full field of view of the capsule endoscope, and the evaluation methods for the resolution effectiveness of pixels number were established. This method selects the horizontal optical axis section as the meridian plane to analyze. The section passing through the planar array sensor is row or column scanned. The line resolution elements number, the line pixel efficiency, the center field matching rate, and the full field maximum matching rate were derived by the analysis unit using the optical resolution angle and the pixel element projection angle. The simple conversion method of resolution length and resolution angle on the spherical field of view can simplify the measurement. These parameters not only constitute the evaluation basis for matching ratio and resolution validity of the pixel number of the capsule endoscope, but also offer a reference for product design, analysis and modification.
  • 加载中
  • [1] 数码相机中的图像传感器和信号处理[M].徐江涛, 高静, 聂海明, 译.北京: 清华大学出版社, 2015.

    Google Scholar

    Nakamura J. Image Sensors and Signal Processing for Digital Still Cameras[M]. Xu J T, Gao J, Nie H M, trans. Beijing: Tsinghua University Press, 2015. Nakamura J.

    Google Scholar

    [2] 王庆有.图像传感器应用技术[M]. 2版.北京:电子工业出版社, 2013.

    Google Scholar

    [3] 陈远.基于垂直层叠结构的CMOS像素和图像传感器研究[D].杭州: 浙江大学, 2009.

    Google Scholar

    Chen Y. Study of CMOS vertical integrated pixel and image sensor[D]. Hangzhou: Zhejiang University, 2009.

    Google Scholar

    [4] 贾晓航, 张沁园, 段晓东, 等.医用内窥镜胶囊式内窥镜: YY 1298-2016[S].北京: 中国标准出版社, 2018.

    Google Scholar

    Jia X H, Zhang Q Y, Duan X D, et al. Medical endoscopes—Capsule endoscopes: YY 1298-2016[S]. Beijing: China Standard Press, 2018.

    Google Scholar

    [5] 贾晓航, 何涛, 颜青来, 等.医用内窥镜硬性内窥镜第1部分: 光学性能及测试方法: YY 0068.1-2008[S].北京: 中国标准出版社, 2009.

    Google Scholar

    Jia X H, He T, Yan Q L, et al. Medical endoscopes—rigid endoscope—Part 1: optical properties and test methods: YY 0068.1-2008[S]. Beijing: China Standard Press, 2009.

    Google Scholar

    [6] 贾晓航, 颜青来, 文燕.医用硬性内窥镜畸变的评定基础和方法[J].光学学报, 2006, 26(8): 1226-1230. doi: 10.3321/j.issn:0253-2239.2006.08.021

    CrossRef Google Scholar

    Jia X H, Yan Q L, Wen Y. Evaluation base and method of medical rigid endoscope distortion[J]. Acta Optica Sinica, 2006, 26(8): 1226-1230. doi: 10.3321/j.issn:0253-2239.2006.08.021

    CrossRef Google Scholar

    [7] 刘茂超, 张雷, 刘沛沛, 等. 300万像素手机镜头设计[J].应用光学, 2008, 29(6): 944-948. doi: 10.3969/j.issn.1002-2082.2008.06.023

    CrossRef Google Scholar

    Liu M C, Zhang L, Liu P P, et al. Design of lens for 3 mega-pixel mobile phone camera[J]. Journal of Applied Optics, 2008, 29(6): 944-948. doi: 10.3969/j.issn.1002-2082.2008.06.023

    CrossRef Google Scholar

    [8] 谷俊达, 向阳.电子内窥镜光学系统设计[J].长春理工大学学报(自然科学版), 2015, 38(2): 18-20, 24. doi: 10.3969/j.issn.1672-9870.2015.02.005

    CrossRef Google Scholar

    Gu J D, Xiang Y. Optical system design of electronic endoscope[J]. Journal of Changchun University of Science and Technology (Natural Science Edition), 2015, 38(2): 18-20, 24. doi: 10.3969/j.issn.1672-9870.2015.02.005

    CrossRef Google Scholar

    [9] 王军华, 卢景红, 徐敏.含有谐衍射面的单片式胶囊内窥镜光学设计[J].激光与光电子学进展, 2012, 49(12): 122203.

    Google Scholar

    Wang J H, Lu J H, Xu M. Optical design of the single-chip capsule endoscopt with harmonic diffraction surface[J]. Laser & Optoelectronics Progress, 2012, 49(12): 122203.

    Google Scholar

    [10] 朱佳巍, 丁桂林.一款超小型广角医用内窥镜镜头的设计[J].激光与光电子学进展, 2014, 51(9): 181-186.

    Google Scholar

    Zhu J W, Ding G L. Design of an ultra-small wide-angle medical endoscope lens[J]. Laser & Optoelectronics Progress, 2014, 51(9): 181-186.

    Google Scholar

    [11] 张艳辉, 黄战华.胶囊内窥镜技术的研究进展[J].现代仪器, 2006, 12(4): 4-7. doi: 10.3969/j.issn.1672-7916.2006.04.002

    CrossRef Google Scholar

    Zhang Y H, Huang Z H. The progress of capsule endoscope technology[J]. Modern Instruments, 2006, 12(4): 4-7. doi: 10.3969/j.issn.1672-7916.2006.04.002

    CrossRef Google Scholar

    [12] 吕联荣, 姜道连, 田春苗.电视原理及其应用技术[M].天津:天津大学出版社, 2001.

    Google Scholar

    [13] 郭鑫, 张薇, 速晋辉, 等.可调焦胶囊内窥镜光学系统设计[J].光子学报, 2015, 44(5): 0522004.

    Google Scholar

    Guo X, Zhang W, Su J H, et al. Design of a focus-tunable capsule endoscope system[J]. Acta Photonica Sinica, 2015, 44(5): 0522004.

    Google Scholar

    [14] 叶斌, 王立强, 石岩, 等.高清医用电子内窥镜微型CMOS摄像模组(英文)[J].光子学报, 2010, 39(11): 1951-1955.

    Google Scholar

    Ye B, Wang L Q, Shi Y, et al. High resolution miniaturized CMOS camera module for medical electronic endoscope[J]. Acta Photonica Sinica, 2010, 39(11): 1951-1955.

    Google Scholar

    [15] 丁琴, 王惠南.采用GRIN透镜的微胶囊内窥成像技术[J].光子学报, 2004, 33(7): 889-892.

    Google Scholar

    Ding Q, Wang H N. An imaging technique of microcapsule endoscope using a GRIN lens[J]. Acta Photonica Sinica, 2004, 33(7): 889-892.

    Google Scholar

    [16] Tseng Y C, Han P, Hsu H C, et al. A flexible FOV capsule endoscope design based on compound lens[J]. Journal of Display Technology, 2016, 12(12): 1-7. doi: 10.1109/JDT.2017.2661078

    CrossRef Google Scholar

    [17] Ouyang M, Jeng W D. Design and analysis of radial imaging capsule endoscope (RICE) system[J]. Optics Express, 2011, 19(5): 4369-4383. doi: 10.1364/OE.19.004369

    CrossRef Google Scholar

  • Overview: Based on clinical application, the capsule endoscope is designed as a wide field of view. However, the wide field of view will cause distortion. The distortion is easy to generate the matching inconsistency between the pixel element and the optical resolution element in the full field of view. Then the effectiveness of the pixels has declined. Furthermore, the distortion even causes the mismatching in some positions. The small intestine capsule endoscope is used to observe the surrounding field of view. The detection found that Moiré pattern occurred in the small intestine capsule endoscopy. Therefore, researches were carried out to establish the evaluation for the matching level between the optical resolution and the pixel number on the full field of view of the capsule endoscope, and the evaluation methods for the resolution validity of pixel number. This method selects the horizontal optical axis section as the meridian plane to analyze. The section passing through the planar array sensor is row or column scanned. The analytical method of the optical resolution angle continuation arrangement was adopted. Then the algorithm of line resolution element number was derived. The number of effective pixels of the line was described by the number of resolution elements. Then, the effective rate of line pixel number was obtained. The pixel element projection angle and the optical resolution angle were taken as the analytical units. The ratio between the pixel element projection angle and the optical resolution angle in the same field of view position was regarded as the line-matching ratio. Three basic functions for analysis were further extended, including the resolution angle density function, the relative function of line matching ratio and the density eigenfunction of pixel element projection angle. A simple measurement method was derived by using the simple conversion relations between a line segment and an included angle on a spherical field of view. Those three basic functions for analysis were obtained by fitting, so that the situation of the pixel element and the optical resolution element in every position, the matching details between two elements and the matching level between two elements in the full field of view were obtained. The parameters, including the line resolution element number, effective rate of line pixel number, center field of view matching ratio, maximum matching ratio of the full field of view and its position and other parameters, not only constitute the evaluation basis for matching ratio and resolution validity of the pixel number of the capsule endoscope, but also offer a reference for product design, analysis and modification.

  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Figures(5)

Tables(2)

Article Metrics

Article views(8150) PDF downloads(2299) Cited by(0)

Access History
Article Contents

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint